top of page

New                                                          
The Global Color Systems GC361T
GCS-361T Transmission Densitometer

Direct replacement for the X-Rite 361T
Calibrated and Designed to align with the X-Rite 361T

The Global Color Systems GC361T
📊 Core Measurement Specs

  • Density range: 0.0 to > 6.0 D

  • Dot area: 0% to 100% (± dot)

  • Measurement capability: Density + dot area on film (positive & negative)

  • Responses: X-Rite Ortho, Visual, and UV

🎯 Accuracy / Repeatability

Repeatability

  • Ortho/Visual (2–3 mm aperture):

    • ±0.01 D (0.0–5.0 D)

    • ±2% (5.0–5.5 D), ±3% (5.5–6.0 D)

  • Ortho/Visual (1 mm aperture):

    • ±0.01 D (0.0–4.5 D)

    • ±1% (4.5–5.0 D)

  • UV (3 mm aperture):

    • ±0.01 D (0.0–3.5 D)

    • ±1% (3.5–4.0 D)

Linearity

  • ±0.02 D (low densities), increasing to ±1–2% at higher densities

Zero stability

  • ±0.02 D 

26-03-18 GCS 361 Pic 2_edited.jpg
bottom of page